Customer is requesting to view polygonal shutters in Universal Viewer . In UV though the dicom info has polygonal shutter info it does not display polygonal shutter, you will see an image with a glaring white surround. Customer complains this is unacceptable and is a visual impairment. Same images can be viewed on RA1000 or WebPACS with black outside the shutter. Customer is on UV 7.0 UV does not support polygonal shutters. |
Example pics from Case 09435384
University Medical Center Management Corporation DBA University Medical Center
Modalities: CR, DX
Site moved from RA1000 to UV v7 SP1.0.1 and is also requesting this RA1000 parity.
Case 09435384
University Medical Center Management Corporation DBA University Medical Center
Modalities: CR, DX
Site moved from RA1000 to UV v7 SP1.0.1 and is also requesting this RA1000 parity.
DICOM Tags that Modality is providing that RA1000 is able to display but UV is not:
(6000,0010) US OLY Rows 3233
(6000,0011) US OLY Columns 1296
(6000,0040) CS OLY Type G
(6000,0050) SS OLY Origin 1\1
(6000,0100) US OLY Overlay Bits Allocated 1
(6000,0102) US OLY Overlay Bit Position 0
(6000,3000) OW OLY Data